| Model Number |
7151H |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The patient/family was the initial reporter, so personal information was not entered.No information was captured as the customer's weight was not provided.This event is related to mdr # 1810909-2020-00413.
|
| |
|
Event or Problem Description
|
|
The customer reported that she obtained a blood glucose reading of 152 mg/dl at 5:47 a.M.On the contour meter.The meter automatically marked it as a control test, and so the reading will be displayed as a control result when accessing the meter's memory.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.Replacement meter kits and test strips were sent to the customer.Since the strip information was not provided, this report will be submitted under the meter information.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer returned the suspected contour meter for evaluation.The customer also returned the bottle of test strips, however, the returned strips were not implicated to be involved in the event occurrence and had already expired.Therefore, the returned meter was tested with the in-house contour test strips using a blood sample, which gave a satisfactory performance.All blood glucose readings obtained during in-house testing were properly marked as blood tests in the meter's memory.
|
| |
|