COVIDIEN DOBBHOFF 8FR;43IN W/STYLTY SI; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884710859E
Medical Device Problem Codes Complete Blockage (1094); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/13/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The complainant indicated that the device will not be returned for evaluation; therefore, a failure analysis is not available, and we are not able to determine the relationship between this device and the cause for this event.If additional information or the sample is received, the investigation will be reopened and responded to accordingly.
 
Event or Problem Description
The customer reports that the enfit connection is clogging and leaking after being in place for seven days.No patient injury.
 
Additional Manufacturer Narrative
Investigation summary: a decontaminated sample was received at the plant for the investigation.Because a lot number was not provided, the device history record could not be performed.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.Upon a visual evaluation of the sample, an occlusion was confirmed, a white mass with a strong smell of medicine was found at the end of the tube; however, there were no leak conditions found in the sample.An investigation was conducted with the multifunctional team.All process and controls were found properly followed, including sub-assemblies, finished product assembly, packaging and inspections performed to the product.There were no abnormal conditions found in our process that could trigger the reported condition.Based on the available information, the most likely root cause indicates that this issue could occur due to inadequate use of the procedure if the instructions for use (ifu) are not followed.Feeding tubes should be flushed frequently to prevent clogging.Suggested flushing schedule is: a) before and after each feeding b) before and after administering medication c) once every four hours during continuous feeding or between intermittent feedings d) each time the feeding set is disconnected e) each time the feeding container is filled/changed f) each time the pump is stopped.The ifu states to use only tap or sterile water to flush.Do not use solutions containing meat tenderizer to flush or open a clogged feeding tube.Administration of medications should be guided by hospital policy.Many liquid preparations contain sorbitol which tends to interact with enteral formulas and clog the feeding tube.Thoroughly crush tablets, excluding enteric tablets which should never be crushed; however, always consult with your pharmacist regarding which tablets should be crushed for feeding tube administration.No action plan is deemed necessary at this time since the reported condition was not confirmed to be manufacturing related.The current process is running according to product specifications meeting quality acceptance criteria.We will keep monitoring the process for any adverse trends that require immediate attention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOBBHOFF 8FR;43IN W/STYLTY SI
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
COVIDIEN
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX  22500
MDR Report Key10413717
Report Number9612030-2020-02568
Device Sequence Number14420262
Product Code KNT
UDI-Device Identifier10884521582880
UDI-Public10884521582880
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number8884710859E
Device Catalogue Number8884710859E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2020
Initial Date Received by Manufacturer 08/13/2020
Supplement Date Received by Manufacturer08/13/2020
Initial Report FDA Received Date08/17/2020
Supplement Report FDA Received Date03/04/2021
Patient Sequence Number1
-
-