Model Number WCD 4000
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/12/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of monitor sn (b)(4) has been completed.The reported problem (service code 105) has been confirmed.Upon investigation the monitor failed a pulse test.The cause for the failure was isolated to shorted insulated-gate bipolar transistors (igbts) q6, q7, q8, q10, q12, and q16 on the defibrillator pca.The root cause for the shorted igbts could not be positively identified.No adverse event resulted from the defective monitor.
 
Event Description
A us distributor reported that a patient's monitor was displaying a service code 105 (pulse test relay stuck).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh, pa
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh, pa
Manufacturer Contact
matthew saich
121 gamma drive
pittsburgh, pa 
9683333
MDR Report Key10414066
MDR Text Key203716679
Report Number3008642652-2020-06652
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005005
UDI-Public00855778005005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 07/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Device Catalogue Number10A0988
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2020
Initial Date Manufacturer Received 07/24/2020
Initial Date FDA Received08/17/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/29/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-