• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. TIGHTROPE ABS, 14 MM BUTTON; PIN, FIXATION, SMOOTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. TIGHTROPE ABS, 14 MM BUTTON; PIN, FIXATION, SMOOTH Back to Search Results
Model Number TIGHTROPE ABS, 14 MM BUTTON
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/29/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
It was reported that a patient had undergone an acl reconstruction (b)(6) 2020.At a post op visit the ar-1588tb-1, tightrope abs 14mm button (lot 10727270), was observed to be broken in half.The patient underwent a second surgery (b)(6) 2020 at the same facility, by the same surgeon.During the second surgery the acl button was removed and an interference screw was utilized to fixate the acl allograft.It was reported that the patient did not suffer any known or reported trauma post the original procedure.
 
Additional Manufacturer Narrative
Complaint confirmed.One unpackaged ar-1588tb-1 was received for investigation.Visual inspection identified that the returned ar-1588tb-1 had split down the centerline.No fragments from the breakage site were returned for investigation.Material analysis concluded that the ar-1588tb-1 button met all as-received specifications.It was noted that the button broke post op, resulting in the revision.As such, potential causes for the reported event include insufficient fixation of the device, and/or non-compliance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIGHTROPE ABS, 14 MM BUTTON
Common Device Name
PIN, FIXATION, SMOOTH
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key10414283
Report Number1220246-2020-02036
Device Sequence Number2682798
Product Code HTY
UDI-Device Identifier00888867019638
UDI-Public00888867019638
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K110123
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model NumberTIGHTROPE ABS, 14 MM BUTTON
Device Catalogue NumberAR-1588TB-1
Device Lot Number10727270
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 08/03/2020
Supplement Date Received by Manufacturer08/03/2020
Initial Report FDA Received Date08/17/2020
Supplement Report FDA Received Date08/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-