• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problem Charging Problem (2892)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/28/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of battery charger/ modem sn (b)(4) has been completed.The reported problem (charger faults) has been confirmed.Upon investigation the battery charger/ modem was unable to charge a battery pack.The cause for the failure was contamination on the battery pca and a shorted q1 current-controlling transistor on the bedside pca.The shorted transistor was caused by the contamination.The root cause for the contamination was ingress of an unknown liquid.No adverse event resulted from the defective equipment.
 
Event Description
A us distributor reported that a patient was experiencing charger faults.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh, pa
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh, pa
Manufacturer Contact
eliza schally
121 gamma drive
pittsburgh, pa 
9683333
MDR Report Key10416552
MDR Text Key203810786
Report Number3008642652-2020-06303
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005029
UDI-Public00855778005029
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 07/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Device Catalogue Number10A0985
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2020
Date Manufacturer Received07/20/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/02/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-