| Model Number |
657324S |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon completion of investigation, additional information will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "the hexlobular internal was not made on screw head and the screw could not be used.The new one was used for the procedure.".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed.The device was returned, and indeed, the torx connection on the screw head was not properly manufactured, making the device unusable.A review of the device history for the reported lot did not indicate any abnormalities.Based on investigation, the root cause was attributed to be manufacturing related.A nonconformity report has been opened in order to further investigate and address this issue for further details.
|
| |
|
Event or Problem Description
|
|
As reported: "the hexlobular internal was not made on screw head and the screw could not be used.The new one was used for the procedure.".
|
| |
|