ABBOTT LABORATORIES (IRVING IA/CC) CARBON DIOXIDE Back to Search Results
Model Number 3L80-21
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/21/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient identifier multiple = (b)(6).There was no additional patient information provided due to privacy issues.This issue was previously reported under mdr number 1415939-2020-00089 under an incorrect manufacturing site: (b)(4).A review of tickets found no other complaints similar to the current complaint for lot 56397uq12.Trending review determined no trends for falsely elevated results for the product.Return testing was not completed as returns were not available.A review of the manufacturing documentation did not identify any issues associated with the customer's observation.A review of the product labeling concluded that the issue is sufficiently addressed.Based on the investigation, no systemic issue or deficiency of carbon dioxide (co2) reagent, lot 56397uq12 was identified.
 
Event or Problem Description
The customer reported falsely elevated bicarbonate (co2) results for the past month on patient samples on the architect analyzer.The customer provided two examples of results: on (b)(6) 2020 (b)(6) initial = 40mmol/l / retest twice =20mmol/l (normal range 22-29mmol/l).There was no reported impact to patient management.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARBON DIOXIDE
Common Device Name
CARBON DIOXIDE
Manufacturer (Section D)
ABBOTT LABORATORIES (IRVING IA/CC)
1915 hurd drive
irving TX 75038
MDR Report Key10417521
Report Number3016438761-2020-00181
Device Sequence Number11365231
Product Code KHS
UDI-Device Identifier00380740005955
UDI-Public00380740005955
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K060295
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2021
Device Model Number3L80-21
Device Catalogue Number03L80-21
Device Lot Number56397UQ12
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/06/2020
Initial Report FDA Received Date08/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-