Model Number G50622
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) #: k163018.Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
 
Event or Problem Description
After they place whole device into the place and they try to deploy the device and they cannot deploy, so they change a new one.
 
Event or Problem Description
Supplemental report being submitted due to the investigation being completed on 16-mar-2021.After they place whole device into the place and they try to deploy the device and they cannot deploy, so they change a new one.
 
Additional Manufacturer Narrative
Pma/510(k) #: k163018.Annex g: g04044 - device deployer.Device evaluation: the zilbs-635-6-8 device of lot number c1520365 involved in this complaint was not returned for evaluation.With the information provided, a document based investigation was conducted.Lab evaluation: n/a.Document review: prior to distribution zilbs-635-6-8 devices are subjected to a visual inspection and functional checks to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the relevant manufacturing records (c1520365) revealed no discrepancies that could have contributed to this complaint.Upon review of complaints, this failure mode has not occurred previously with this lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1520365.There is no evidence to suggest that the customer did not follow the instructions for use (ifu0065-3).Root cause review: a definitive root cause could not be determined from the available information.A possible root cause could be attributed to tortuous anatomy which may have compressed or rotated the delivery system as it advanced and prevented the stent from being deployed.Summary: the complaint is confirmed based on customer testimony.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZILVER 635 BILIARY SELF EXPANDING METAL STENT
Common Device Name
FGE CATHETER, BILIARY, DIAGNOSTIC - BILIARY STENT - METAL
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key10417662
Report Number3001845648-2020-00504
Device Sequence Number2942730
Product Code FGE
UDI-Device Identifier10827002506227
UDI-Public(01)10827002506227(17)210718(10)C1520365
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/18/2021
Device Model NumberG50622
Device Catalogue NumberZILBS-635-6-8
Device Lot NumberC1520365
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date07/20/2020
Event Location Hospital
Initial Date Received by Manufacturer 07/22/2020
Supplement Date Received by Manufacturer07/22/2020
Initial Report FDA Received Date08/17/2020
Supplement Report FDA Received Date03/31/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age66 YR
Patient Weight60
-
-