| Model Number |
1012632-19 |
| Medical Device Problem Codes |
Difficult or Delayed Activation (2577); Migration (4003)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
07/24/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product was not returned to abbott vascular for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.The investigation determined the reported difficulties appear to be related to operational circumstances of the procedure.Based on the reported information, it is likely that the anatomical conditions prevented the stent from fully deploying.Manipulation of the device likely caused the stent to migrate.There is no indication of a product quality issue with respect to the design, manufacture.
|
| |
|
Event or Problem Description
|
|
It was reported the procedure was to treat a target lesion in the common iliac artery.When deploying the 9.0mmx19mmx135cm omnilink stent in the target lesion, there was lack of distal deployment on the distal edge of the stent and the stent migrated into healthy tissue in the external iliac artery.The delivery system was withdrawn to the migrated stent and the stent was implanted in healthy tissue and fully apposed to the vessel wall.A second omnilink stent was implanted without issue in the target lesion.No additional treatment was performed.No additional information was provided.
|
| |
|