| Model Number |
38270 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: returned product consisted of an imager diagnostic catheter in the un-opened sterile pouch.Tensile strength inspection of the returned product was performed and the device failed to meet specification for tip detachment.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 23jul2020.A 5f bern imager ii angiographic catheter was returned to bsc from a customer with no reported complaint against the device.Device analysis on the returned device found that the device failed tensile testing leading to a tip detachment.
|
| |
|