Model Number SDK4000-FT
Medical Device Problem Codes Device Displays Incorrect Message (2591); Application Program Problem (2880); Patient Data Problem (3197)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 07/23/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during the procedure, the physician could not get an adequate registration in order to advance to navigation.Something seemed to be interfering with the magnetic field.Did all of the proper trouble shooting and continued to have the same failed result.Registration review never showed the data points within the 3d tree.Finally, a nurse mentioned that the patient had lots of metal in the neck and spine.Did a fluoro shot and confirmed that there were multiple metal rods in the patient's neck and upper spine.The procedure was not completed and was not completed by other means.The patient was under general anesthesia.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDGE
Common Device Name
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
COVIDIEN LP - SUPERDIMENSION INC
161 cheshire lane, suite 100
plymouth MN 55441
MDR Report Key10417993
Report Number3004962788-2020-00048
Device Sequence Number17137512
Product Code JAK
UDI-Device Identifier10884521545557
UDI-Public10884521545557
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K151376
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSDK4000-FT
Device Catalogue NumberSDK4000-FT
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2020
Initial Report FDA Received Date08/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-