Model Number CQ7588
Medical Device Problem Codes Deflation Problem (1149); Material Frayed (1262); Peeled/Delaminated (1454)
Health Effect - Clinical Code Injury (2348)
Date of Event 07/27/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 08/2022).
 
Event or Problem Description
It was reported that during an angioplasty procedure, the balloon allegedly failed to deflate.It was further reported that a percutaneous needle stick was used to deflate the balloon.The patient status was normal.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the balloon allegedly failed to deflate.It was further reported that a percutaneous needle stick was used to deflate the balloon.The patient status was normal.
 
Additional Manufacturer Narrative
H10: manufacturing review: a lot history review was conducted and it was determined that a device history record review was not required.Investigation summary: the device was returned for evaluation.A visual inspection found peeled pebax and frayed fibers on the outside of the balloon.An attempt was made to inflate the balloon, however a compound rupture was noted on the balloon.Therefore, the investigation is confirmed for peeling pebax, frayed fibers, and for a compound rupture.However, the investigation is inconclusive for the reported deflation issue as the balloon was unable to be inflated to functionally test for the deflation issue.It is possible that the identified balloon rupture could have contributed to the reported deflation issue and identified frayed fibers and pebax peeling.However, the definitive root cause for the identified failures cold not be determined based on the available information.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4(expiry date: 08/2022),.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key10418752
Report Number2020394-2020-05411
Device Sequence Number3190410
Product Code DQY
UDI-Device Identifier00801741063138
UDI-Public(01)00801741063138
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K083657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCQ7588
Device Catalogue NumberCQ7588
Device Lot NumberREDW3546
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2020
Initial Date Received by Manufacturer 07/30/2020
Supplement Date Received by Manufacturer11/17/2020
Initial Report FDA Received Date08/18/2020
Supplement Report FDA Received Date11/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-