WRIGHTS LANE SYNTHES USA PRODUCTS LLC CRANIOS REINFORCED FAST SET PUTTY 10CC-STERILE; METHYL METHACRYLATE FOR CRANIPLASTY Back to Search Results
Model Number 615.10.01S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hearing Impairment (1881); Pain (1994)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.Investigation summary: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history: part number: 615.10.01s, cranio's reinforced fast set putty 10cc ¿ sterile.Lot number: dsd8462 (sterile).Manufacturing location: (b)(4) / inspected and released by: (b)(4).Release to warehouse date: 30-dec-2016.Expiration date: 28-jul-2018.Lot quantity: 126.Purchased finished goods traveler met all inspection acceptance criteria.Inspection sheet, incoming final inspection, met all inspection acceptance criteria.Certificate of conformance supplied by (b)(4) dated 28-dec-2016 was reviewed and determined to be conforming.Sterility parameters documented on certificate were reviewed and gama is specified at 25 min.- 32 max.The max results are documented at 31.9 and 32.2.It is unknown whether exceeding specified maximum may have contributed to the reported complaint condition.This lot met all visual and packaging criteria at the time of release with no issues documented during the inspection or release of the product that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported on an unknown date, the patient underwent an acoustic neuroma resection retrosigmoid procedure.The patient was implanted with the titanium matrixneuro adaptation plate and putty cement.There was a surgical time of 559 minutes.There were no intra-operative complications.The patient is doing great after gross total resection of a massive vestibular schwannoma and a scar in the iac.The patient reports low back pain and issues with sleeping recently.Patient has no hearing on the right side.This is report 1 of 3 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRANIOS REINFORCED FAST SET PUTTY 10CC-STERILE
Common Device Name
METHYL METHACRYLATE FOR CRANIPLASTY
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10420109
Report Number2939274-2020-03588
Device Sequence Number3065910
Product Code GXP
UDI-Device Identifier10887587058658
UDI-Public(01)10887587058658
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K102018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source study
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/28/2018
Device Model Number615.10.01S
Device Catalogue Number615.10.01S
Device Lot NumberDSD8462
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/06/2020
Initial Report FDA Received Date08/18/2020
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MATRIXNEURO ADAPT-PL 20HO T0.4 TI; UNK - SCREWS: TRAUMA
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR
-
-