• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC FUSION COMPACT; EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC FUSION COMPACT; EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735602
Device Problem Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/13/2020
Event Type  malfunction  
Manufacturer Narrative
No products have been returned to medtronic for analysis.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system used prior to a functional endoscopic sinus surgery (fess) procedure.It was reported that when the system was booted up, a red x showed on the emitter tracking details.The site rebooted the system and reseated the emitter with no resolution.Technical services (ts) then talked them through shutting down the system and powering on the axiem box and monitor separately then reconnecting.The issue was resolved.There was a reported delay to the procedure of thirty minutes.There was no reported impact to the patient.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION COMPACT
Type of Device
EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
tricha miles
7000 central avenue ne rcw215
minneapolis, MN 55432
7635140379
MDR Report Key10420341
MDR Text Key203343270
Report Number1723170-2020-02200
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153247
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735602
Device Catalogue Number9735602
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/10/2020
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age48 YR
Patient Weight104
-
-