• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL DELTAXSFT10 3MM X 6CM; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL DELTAXSFT10 3MM X 6CM; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number DLX100306
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Procode: krd/hcg.Initial reporter: the customer contact information, including name, occupation, phone, fax, and e-mail address, was not reported.The product is available for evaluation and testing; however, it has not been received to date (which is indicated as "other").Additional information will be submitted within 30 days of receipt.The company is seeking this information through the event investigation.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Event or Problem Description
As reported by the filed, during a coil embolization, the physician didn't like the shape of the coil deployment of a 3mm x 6cm (dlx100306, l15976) and decided to pull it out.During resheathing of the coil, the introducer failed to be re-zipped.The physician used a same like product.The procedure was successfully finished.There was not patient injury reported.This coil was being used for framing.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.Complaint conclusion: as reported by the filed, during a coil embolization, the physician didn't like the shape of the coil deployment of a 3mm x 6cm (dlx100306, l15976) and decided to pull it out.During resheathing of the coil, the introducer failed to be re-zipped.The physician used a same like product.The procedure was successfully finished.There was not patient injury reported.This coil was being used for framing.A non-sterile unit deltaxsft10 3mm x 6cm was received inside of a pouch.The device was inspected, and it was found that the introducer was kinked at 85.5 cm from proximal, the re-sheating tool and the hub was not returned with the rest of the device.Also, the marker band was found at 40.7 cm from hub, and it was found within specification.The embolic coil was inspected under microscope and it was found in good normal conditions.The functional analysis could not be performed due to the returned condition of the device (missing re-sheating tool).The complaint reported by the customer ¿coil introducer - zipping difficulty-rezipping¿ could not be duplicated since the device was returned without the re-sheating tool, also the kinked condition noted on the dpu could be related with the complaint reported.Due those conditions, the complaint reported by the consumer cannot be confirmed.The complaint reported by the customer ¿coil - positioning difficulty-poor conformability¿ could not be evaluated because the complaint reported is too specific to the patient and procedure at the time of occurrence and cannot be replicated in the lab.Also, the embolic coil was inspected, and it was found in good normal conditions.The kinked condition noted on the dpu and the missing re-sheating tool could be related with the costumer complaint regarding zipping difficulty.Those conditions may have appeared due to excessive force and handling being applied to the device however none of those can be conclusively determined.Neither the analysis nor the mre suggest that the failure reported by the costumer could be related to the manufacturing process.The instructions for use (ifu) instructs the user to unsheathe a small length of the dpu to unlock the device, then to advance the embolic coil into the microcatheter until the hub connector of the dpu reaches the proximal end of the resheathing tool.This results in the placement of the embolic coil and the more flexible and fragile distal sections of the dpu inside the microcatheter before continuing to unsheathe the device.If the device is unsheathed before advancing into the microcatheter, there is a risk that the embolic coil or the distal end of the dpu will be unsheathed and pass through the resheathing tool.If the resheathing tool passes over the distal end of the dpu, this will expose these flexible sections, which will then be subject to kinking and bending.Once a part of the dpu is kinked or bent, the translucent introducer sheath will not be able to re-form around it, and that section of the device will protrude.The ifu instructs on proper positioning of the microcoil system.The ifu also warns: ¿if repositioning of the microcoil is necessary, carefully observe the motion of the microcoil in respect to the dpu wire while retracting the microcoil under fluoroscopy.If the microcoil movement is not one-to-one with the dpu wire, or if repositioning is difficult, the microcoil may have become stretched and could possibly break.Gently remove and discard the microcoil system¿.Assignment of root cause for the events remains speculative and inconclusive, based on the limited information provided and the evidence presented by the returned device; however, it is possible that clinical and procedural factors, including device manipulation and device interaction, may have contributed to the reported failures and damages on the returned system.Multiple attempts to obtain additional information were unsuccessful.If information is provided at a later date, the file will be reopened and processed accordingly.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the events were related to a manufacturing or design issue, no corrective actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTAXSFT10 3MM X 6CM
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
MDR Report Key10420345
Report Number3008114965-2020-00363
Device Sequence Number4060058
Product Code KRD
UDI-Device Identifier10886704077350
UDI-Public10886704077350
Combination Product (Y/N)N
PMA/510(K) Number
K150319
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Model NumberDLX100306
Device Catalogue NumberDLX100306
Device Lot NumberL15976
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2020
Initial Date Received by Manufacturer 07/30/2020
Supplement Date Received by Manufacturer09/01/2020
Initial Report FDA Received Date08/18/2020
Supplement Report FDA Received Date09/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-