| Model Number |
765100 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
|
| |
|
Event or Problem Description
|
|
The customer reported the device leaks from the base of the bag.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record shows evidence that the product was released according to all established procedures and qa documentation.Eighteen brand new devices were received for evaluation.A visual and functional inspection found no evidence of a leak in any of the devices.The root cause cannot be identified as the reported condition could not be confirmed based on the test results for the returned samples.A formal corrective/preventative action will not be initiated at this time as the reported condition did not identify a manufacturing or production process.This complaint will be closed with no further action and will be used for tracking and trending purposes.
|
| |
|