Model Number C8XX2
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event unit returned to applied medical for evaluation.A follow-up report will be provided upon completion of the investigation.
 
Event or Problem Description
Name of procedure being performed: left hand-assisted nephrectomy.Detailed description of event: during the case, the alexis sheath was first deployed within the incision.The gel cap was placed over the alexis and a 5mm [name] trocar was used through the cap for about 30 minutes.At that point the procedure was switched to open and the gel seal cap was removed.When the cap was removed, the sheath fell away from the ring and the alexis was no longer able to provide retraction at the incision site.The device did not fragment.The device was removed from the patient and a new alexis was opened to provide retraction and complete the case.No patient injury.The device is available for return.Additional information received via email on 03aug2020 from [name]: photos have been provided of the device packaging.Additional information received via email on 12aug2020 from [name]: "unfortunately, my contact [name] general service lead did not remember which ring the sheath separated from.The only picture that she took was the outside of the box same one that i provided to you with the lot number.I did confirm the inner ring was able to be pulled out of incision like it typically would.No patient harm and no fragmentation." patient status: no patient injury.Type of intervention: the device was removed from the patient and a new device was opened to provide retraction and complete the case.
 
Additional Manufacturer Narrative
The event unit was returned to applied medical for evaluation.Upon visual inspection, engineering confirmed that the sheath had separated from the inner ring.Based on the condition of the returned unit, it is likely the reported event was caused by a weak or incomplete heat seal between the sheath and the inner ring.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
 
Event or Problem Description
Name of procedure being performed: left hand-assisted nephrectomy.Detailed description of event: during the case, the alexis sheath was first deployed within the incision.The gel cap was placed over the alexis and a 5mm [name] trocar was used through the cap for about 30 minutes.At that point the procedure was switched to open and the gel seal cap was removed.When the cap was removed, the sheath fell away from the ring and the alexis was no longer able to provide retraction at the incision site.The device did not fragment.The device was removed from the patient and a new alexis was opened to provide retraction and complete the case.No patient injury.The device is available for return.Additional information received via email on 03aug2020 from [name] : photos have been provided of the device packaging.Additional information received via email on 12aug2020 from [name]: "unfortunately, my contact [name] general service lead did not remember which ring the sheath separated from.The only picture that she took was the outside of the box same one that i provided to you with the lot number.I did confirm the inner ring was able to be pulled out of incision like it typically would.No patient harm and no fragmentation." patient status: no patient injury.Type of intervention: the device was removed from the patient and a new device was opened to provide retraction and complete the case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C8XX2, GELPORT 120MM
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
MDR Report Key10420834
Report Number2027111-2020-00530
Device Sequence Number16920306
Product Code GCJ
UDI-Device Identifier20607915123912
UDI-Public(01)20607915123912(17)211110(30)01(10)1384688
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K014047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/10/2021
Device Model NumberC8XX2
Device Catalogue Number700314608
Device Lot Number1384688
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/23/2020
Supplement Date Received by Manufacturer07/23/2020
Initial Report FDA Received Date08/18/2020
Supplement Report FDA Received Date08/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5MM COVIDIEN TROCAR
-
-