RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 18aug2020.
|
|
Event Description
|
It was reported that the unit failed ground resistance at the oxygen fitting.The customer replaced the flow sensor assembly.While running performance verification testing (pvt) the unit failed ground resistance at the oxygen fitting.The remote service engineer (rse) advised to take the locktite off.The customer reported the unit returning to service.No parts were replaced.The unit was not in use, and there was no patient or user harm reported.
|
|