• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL LOCKING SCREW, FULLY THREADED T2 TIBIA 5X45 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL LOCKING SCREW, FULLY THREADED T2 TIBIA 5X45 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 1896-5045S
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Failure of Implant (1924); Injury (2348); Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 05/26/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Upon completion of investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
The customer reported a broken gamma nail which is in need of revision.Revision surgery was scheduled for (b)(6) 2020.Additionally reported: the surgeon commented that the patient was a smoker and other patient factors had led to delayed union and that¿s why he felt implant had broken.Locking screw was also broken.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the device was returned for evaluation and matches the alleged failure mode.Device inspection revealed the following: the received locking screw was found to be broken around the distal region.The tip of the screw had sheared off.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the above investigation, the root cause of the failure is patient related (patient factors issue).The surgeon commented that the patient was a smoker and other patient factors had led to delayed union and that¿s why he felt implant had broken.Smoking cigarettes has been shown to have a significant impact on broken bone healing.The labeling also states ¿these devices can break when subjected to the increased loading associated with delayed unions and/or non-unions.Internal fixation devices are load sharing devices which are intended to hold fractured bone surfaces in apposition to facilitate healing.If healing is delayed or does not occur, the appliance may eventually break due to metal fatigue.Loads on the device produced by load bearing and the patient¿s activity level will dictate the longevity of the device.¿.If any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
The customer reported a broken gamma nail which is in need of revision.Revision surgery was scheduled for (b)(6) 2020.Additionally reported: the surgeon commented that the patient was a smoker and other patient factors had led to delayed union and that¿s why he felt implant had broken.Locking screw was also broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW, FULLY THREADED T2 TIBIA 5X45 MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
MDR Report Key10423193
Report Number0009610622-2020-00478
Device Sequence Number11680463
Product Code HSB
UDI-Device Identifier04546540202505
UDI-Public04546540202505
Combination Product (Y/N)N
Initial Reporter CountryGB
PMA/510(K) Number
K032244
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/28/2023
Device Model Number1896-5045S
Device Catalogue Number18965045S
Device Lot NumberK06CEDC
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2020
Initial Date Received by Manufacturer 07/22/2020
Supplement Date Received by Manufacturer09/17/2020
Initial Report FDA Received Date08/18/2020
Supplement Report FDA Received Date10/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-