| Model Number |
BFH150 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Death (1802)
|
| Date of Event |
04/12/2020
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
A second thermometer was returned to our company by the same consumer.The second device was not related to this adverse event, and was a different model.The consumer stated that she believed that the second device was also giving false negative readings.Both devices were found to be within the 0.4°f tolerance for accuracy when tested in our lab.
|
| |
|
Event or Problem Description
|
|
A consumer reported that her thermometer had allegedly given false negative readings on her step-mother.It was alleged that the device gave readings that were 2.5 - 5°f lower than what was later measured in an ambulance.The patient was admitted into a hospital and tested positive for covid-19.She passed away 2 days after being admitted into the hospital.It was alleged that the false negative readings given by the device may have caused a delay in medical treatment, but that she did not feel well for several days prior to calling an ambulance.It was stated that the patient had pre-existing medical conditions which put her at high risk, and also stated that catching the fever earlier with a working thermometer would likely not have made a difference but it would have been nice to know that she had a fever.
|
| |
|