• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. EB EXC GRN 30IN 2-0 S/A SH; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. EB EXC GRN 30IN 2-0 S/A SH; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Model Number X833H
Medical Device Problem Codes Material Frayed (1262); Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/21/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested and the following was obtained: was there any adverse consequence associated with the patient? there were no adverse consequences associated with the patient.
 
Event or Problem Description
It was reported that a patient underwent a bariatric procedure on (b)(6) 2020 and suture was used.During the procedure, the needle easily popped off the suture at the swage.The suture would also easily fray.There were no adverse patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Date sent to the fda: 09/17/2020.A manufacturing record evaluation was performed for the finished device batch, and no non-conformances were identified.H-3 summary: an opened box with thirty-two unopened samples of product were returned for analysis.The complaint sample was not received for evaluation.During the visual inspection of samples, no defects were found on the packages.The samples were opened, and swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, fraying or damaged were observed.A functional test was performed using an instron and the pull force were above the minimum requirements.Per the conditions of the samples received, no performance pull off suture needle or fraying suture was found, and the tested samples met the finished goods requirements.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EB EXC GRN 30IN 2-0 S/A SH
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876
MDR Report Key10423686
Report Number2210968-2020-06193
Device Sequence Number2961201
Product Code GAT
UDI-Device Identifier10705031058033
UDI-Public10705031058033
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model NumberX833H
Device Catalogue NumberX833H
Device Lot NumberPMH788
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2020
Initial Date Received by Manufacturer 08/10/2020
Supplement Date Received by Manufacturer08/24/2020
Initial Report FDA Received Date08/18/2020
Supplement Report FDA Received Date09/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-