Date sent to the fda: 09/17/2020.A manufacturing record evaluation was performed for the finished device batch, and no non-conformances were identified.H-3 summary: an opened box with thirty-two unopened samples of product were returned for analysis.The complaint sample was not received for evaluation.During the visual inspection of samples, no defects were found on the packages.The samples were opened, and swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, fraying or damaged were observed.A functional test was performed using an instron and the pull force were above the minimum requirements.Per the conditions of the samples received, no performance pull off suture needle or fraying suture was found, and the tested samples met the finished goods requirements.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|