Model Number 75446535
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/27/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation of the returned device was not completed at the time of this report.A follow up report will be sent when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information received from initial reporter via medtronic field representative regarding an event happened during an intra-op for a p re-operative diagnosis of lumbar degenerative disease.It was reported that during the screw implantation process, the tip of the screwdriver was broken in the screw.After the screw was removed, a new screw was implanted and the operation continued.The status of the screw remained as explanted-complete.There were no patient symptoms were reported due to this event.There were no complications to patient/physician were reported/anticipated.Additional information received on 31-jul-2020: the damaged screw was replaced with new screw in the same surgery.
 
Additional Manufacturer Narrative
H3: product analysis #(b)(4):(b)(4) lot# h5483385 analysis found, after visual and optical examination and functional testing, it does not appear to be any damage nor does it indicate any functional issues with the screw.There are some witness marks in the hex area.The screw and the broken hex driver tip where received separated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT PUERTO RICO OPERATIONS CO, MED REL
road 909, km. 0.4., barrio mar
humacao PR 00792
MDR Report Key10424063
Report Number1030489-2020-01096
Device Sequence Number2938663
Product Code KWP
UDI-Device Identifier00885074134109
UDI-Public00885074134109
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K042025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75446535
Device Catalogue Number75446535
Device Lot NumberH5483385
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/27/2020
Supplement Date Received by Manufacturer11/06/2020
Initial Report FDA Received Date08/19/2020
Supplement Report FDA Received Date11/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
-
-