| Model Number |
ANKLE TIBIAL CENTER PUNCH |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/12/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device pending evaluation.
|
| |
|
Event or Problem Description
|
|
As reported, during the punch guide broke during the impaction.No parts or pieces fell into the wound site.There was no impact to the patient.Device will be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
(h3) the broken center peg punch reported may have been the result of both high impaction forces and the patient¿s hard bone.However, this cannot be confirmed as the device was not available for evaluation.
|
| |
|
Search Alerts/Recalls
|