Model Number 765100
Medical Device Problem Codes Fluid/Blood Leak (1250); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/15/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
 
Event or Problem Description
The customer reported the purple cap broke from the tubing causing the formula to leak.There was no patient harm.
 
Additional Manufacturer Narrative
A review of the device history record revealed no discrepancies that may have contributed to the reported condition.All quality assurance testing is performed during the manufacturing of the product met specification requirements.There were 181 devices received for evaluation.A visual and functional inspection was performed and found that many of the cross spikes were detached from the tubing line.The root cause and action plan will be documented through a formal corrective/preventative action which has been initiated to address the reported condition to mitigate any further occurrences.This action is currently ongoing.This complaint will be closed with no further action and will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOEY 1000ML SAFETY SCREW SPIKE
Common Device Name
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
15 hampshire street
mansfield MA 02048
MDR Report Key10425685
Report Number1282497-2020-09361
Device Sequence Number2670900
Product Code LZH
UDI-Device Identifier20884521155784
UDI-Public20884521155784
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number765100
Device Catalogue Number765100
Device Lot Number193290069
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 08/11/2020
Supplement Date Received by Manufacturer08/11/2020
Initial Report FDA Received Date08/19/2020
Supplement Report FDA Received Date01/07/2021
Patient Sequence Number1
-
-