| Catalog Number |
66000016 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, fourth layer of profore 18-25cm ankle circ is not rolling out or malformed (shredded).A backup device was used.Information of the patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the device intended for use in treatment was not returned for evaluation, all provided information has been reviewed and we have not been able to establish a relationship between the reported event or determine a root cause.This investigation has not been able to determine a definitive root cause, however factors that can contribute to the reported issues are temperature fluctuation during transport and storage.If the material is subject to temperature fluctuations, then this can affect the use.The manufacturing records show no evidence that the product did not meet specification at the time of manufacture.A complaint history review found other related events.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
|
| |
|
Search Alerts/Recalls
|