• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ENDOBUTTON CL ULTRA 15MM; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ENDOBUTTON CL ULTRA 15MM; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72200146
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Information (3190)
Date of Event 07/28/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during cl reconstruction surgery, when the tendon traction was being performed, the loop of the endobutton and bone surface were cut and broken.The procedure was completed without significant delay using a back-up device.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H2: additional information on b5 and b7.
 
Event or Problem Description
It was reported that during cl reconstruction surgery, when the tendon traction was being performed, the loop of the endobutton was cut and broken by the bone surface.The bone did not suffer any damage.The procedure was completed without significant delay using a back-up device.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
One 72200146 15mm ultra continuous loop endobutton used in treatment was returned for evaluation.The product fibers were visibly disturbed indication was a difficult use.The loop was unusually tainted, tattered and frayed.Running the blue green suture through the loop had obviously disrupted the fibers.The report indicated damage caused by contact with sharp bone; subsequent removal required.The product fibers are subject to fluffing upon removal from the package.Suture may catch and disrupt the loop fibers if the endobutton loop is not carefully held aside during adjustment.Finesse and care is necessary with handling the product.The nature of the filament fiber is extremely fine and the condition as a perceived flaw.There is no detriment to safety, efficacy or design intent.Complaint history review indicated no similar allegations for the lot number reported.Batch review did not indicate a condition, product or procedure failure that supported the allegation.Instructions for use contains recommendations and precautionary statements for proper use of product.Risk management files contain the reported failure.No root cause related to the manufacture of the device was confirmed.Product met specifications upon release to distribution.Per clinical/mi: no relevant clinical information is provided, recommend closure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOBUTTON CL ULTRA 15MM
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key10426670
Report Number1219602-2020-01281
Device Sequence Number11390021
Product Code MBI
UDI-Device Identifier03596010597250
UDI-Public03596010597250
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K980155
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/05/2022
Device Model Number72200146
Device Catalogue Number72200146
Device Lot Number50657082
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2020
Initial Date Received by Manufacturer 07/29/2020
Supplement Date Received by Manufacturer08/20/2020
10/08/2020
Initial Report FDA Received Date08/19/2020
Supplement Report FDA Received Date08/27/2020
10/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-