It was reported that during the procedure that the trial was chipped, outside the patient.Not any piece fell into the patient and all pieces were recovered.The procedure was completed without delay and backup from smith&nephew was available to finish the surgery.No patient injury or other complication were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
H3, h6: the associated device, used in treatment, was returned, and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device is damaged with chips, rendering the device inoperable.The device was manufactured in 2018 and shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|