ABBOTT VASCULAR XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1074250-38
Medical Device Problem Codes Difficult to Remove (1528); Stretched (1601); Improper or Incorrect Procedure or Method (2017); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/16/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device was received.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
 
Event or Problem Description
It was reported that during removal of the protective sheath from the 2.5x38mm rx xience xpedition drug eluting stent (des), resistance was noted and force was applied.The stent elongated on the balloon and became stuck in the protective sheath.There was no patient involvement and no clinically significant delay in the procedure.Another unspecified device was used to complete the procedure.No additional information was provided.
 
Event or Problem Description
Subsequent to the initial 30-day medwatch report, the following information was provided: the stent remained stuck in the protective sheath and came off of the balloon.
 
Additional Manufacturer Narrative
A visual and dimensional inspection was performed on the returned device.The reported stent dislodgement and stretched stent was confirmed.The reported difficult to remove from the protective sheath could not be tested due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents.The investigation determined the reported difficulties appear to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10427968
Report Number2024168-2020-06929
Device Sequence Number2938908
Product Code NIQ
UDI-Device Identifier08717648162183
UDI-Public08717648162183
Combination Product (Y/N)Y
Initial Reporter CountryCA
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/06/2022
Device Model Number1074250-38
Device Catalogue Number1074250-38
Device Lot Number9093041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2020
Initial Date Received by Manufacturer 07/27/2020
Supplement Date Received by Manufacturer08/28/2020
Initial Report FDA Received Date08/19/2020
Supplement Report FDA Received Date09/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-