• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ANTHOLOGY SO TR NECK SZ 1-6; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ANTHOLOGY SO TR NECK SZ 1-6; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71365701
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/20/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the magnet on the neck trial was found cracked.This was discovered after the case.The patient was not impacted.
 
Additional Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device confirms the magnet on the device is damaged and cracked.The device is manufactured in 2005.The device exhibits signs of significant wear and use.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHOLOGY SO TR NECK SZ 1-6
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10428419
Report Number1020279-2020-04064
Device Sequence Number19306925
Product Code JDG
UDI-Device Identifier03596010559814
UDI-Public03596010559814
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71365701
Device Catalogue Number71365701
Device Lot Number05JM00572
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 07/20/2020
Supplement Date Received by Manufacturer09/16/2020
Initial Report FDA Received Date08/19/2020
Supplement Report FDA Received Date09/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No