A endurant ii bifurcate stent graft system was implanted in the endovascular treatment of a 55mm abdominal aortic aneurysm.It was noted hours after the procedure, during a ct, that the limb etlw1610c124e had occluded.Intervention was performed and the patient had a fem-fem bypass completed.As per the physician the cause of the event was user error.No additional clinical sequelae were provided and the patient is fine.
|
|
In relation to the cause given as user error, it was reported that during the procedure, unfortunately the contralateral limb was deployed outside the contralateral gate.After the physician had thought they had cannulated the gate, the pigtail catheter was rotated in the proximal portion of graft, the contralateral limb was then deployed.All looked normal at the end of the procedure, there was no endoleak and both limbs filled normally.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|