• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PROLENE BLU 30IN 7-0 S/A BV175-8 MP; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PROLENE BLU 30IN 7-0 S/A BV175-8 MP; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE Back to Search Results
Model Number EH7402H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/10/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: how the product was defective? sutures broke at knot.Patient¿s current status? no harm to patient.Please provide procedure date- (b)(6) 2020.
 
Event or Problem Description
It was reported that a patient underwent a vascular procedure on (b)(6) 2020 and suture was used.During the procedure, the suture broke.There were no adverse patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Date sent to the fda: 10/09/2020.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 evaluation: representative samples returned to support investigation of multiple device issues captured in reports 2210968-2020-06244, 2210968-2020-06245, 2210968-2020-06246, 2210968-2020-06247, 2210968-2020-06248, and 2210968-2020-06249.Analysis results have been included in follow-up reports for each.Two sealed boxes and unopened samples of product were received for evaluation.During the visual inspection of unopened samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand with no defects, damage or suture breakage noted.A functional test was performed and the tensile strength force was above the minimum requirement.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.The device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.The device performed without any defect noted and the actual complaint sample was not returned for analysis.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc., or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROLENE BLU 30IN 7-0 S/A BV175-8 MP
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876
MDR Report Key10431199
Report Number2210968-2020-06245
Device Sequence Number1875029
Product Code GAW
UDI-Device Identifier10705031096301
UDI-Public10705031096301
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2024
Device Model NumberEH7402H
Device Catalogue NumberEH7402H
Device Lot NumberPPBCQ1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2020
Initial Date Received by Manufacturer 08/13/2020
Supplement Date Received by Manufacturer09/18/2020
Initial Report FDA Received Date08/20/2020
Supplement Report FDA Received Date10/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-