Model Number PFSDV0016
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/27/2020
Event Type  malfunction  
Event Description
It was reported that during a navio ukr procedure, after implant planning when the surgeon wanted to burr the femur, the message handpiece exposure control motor failure appeared.They opened another navio set instrument to swap the handpiece to continue the procedure without issues.There was a delay of 10 minutes.After the surgery, both handpiece and anspach were tested and were fine.No other complications were reported.
 
Manufacturer Narrative
The navio handpiece p/n 110004, s/n (b)(6) used for treatment was returned for evaluation.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.The reported problem was confirmed.The "could not communicate with handpiece error" appeared at the start of the handpiece test.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.The most likely cause of this event is mishandling during shipping, storage, use or reuse of the device, wear and tear over time.The use of care and caution should be exercised during the surgical site setup and tear down to protect the drill and handpiece cable from sharp objects or from situations that pin the cable between two objects.Do not use excessive force on the drill and handpiece strain reliefs during the decontamination process to minimize separating the cable from the strain relief.Refer to the navio surgical system user's manual and the navio surgical system instrument kit cleaning and sterilization guide for proper handling.Based on the investigation, no further containment or corrective action is recommended or required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO HANDPIECE
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key10432027
MDR Text Key204221875
Report Number3010266064-2020-01723
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556628904
UDI-Public00885556628904
Combination Product (y/n)N
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/20/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPFSDV0016
Device Catalogue NumberPFSR110137
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2020
Date Manufacturer Received06/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CN: NPFS02020, SN: (B)(4)
-
-