| Model Number |
26159BE |
| Medical Device Problem Codes |
Material Twisted/Bent (2981); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Per infrastructure's evaluation of the device: the electrode is bent.In the bent area on the outside of the curve, the insulation is shaved off; remaining chips are visible in the damaged area.Conclusion: bent electrode got shaved off--most likely on the edge of a trocar.
|
| |
|
Event or Problem Description
|
|
As per a manufacturer incident report we received from the factory in (b)(6): when the doctor is operating with 26159be bipolar dissecting electrode, middle part of the electrode insulator came off during the procedure, resulting in foreign body in patient's body.Part got retrieved - no harm to patient.
|
| |
|