(b)(4).Investigation summary: level b investigation.Complaint evaluation / complaint history check for the event(s) that occurred.Severity: s_2__; occurrence: a complaint history check was performed and this is the 1st related complaint for does not detach as intended, needle hub separates and needle stick (clean) on lot # 9266921.A review of risk management 150rmn-0001-16 revision 13 indicates that the potential risk of this specific reported incident (syringe, shield does not detach as intended, needle hub separates, needle stick) was captured and addressed investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed.If samples are received in the future the complaint will be reopened for further investigation.A review of the device history record was completed for batch # 9266921 all inspections were performed per the applicable operations qc specifications.There was one (1) notification [200879125] noted that did not pertain to the complaint.Investigation conclusion: as no samples and/or photo(s) were received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Root cause description: root cause cannot be determined at this time as the issue is unconfirmed as no samples or photos were returned.Rationale: based on the investigation, no additional investigation and no capa is required at this time.
|