Investigation: the following allegations have been investigated: vena cava (vc) perforation, tilt, chest/leg/pain, limited mobility.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported chest/leg/pain and limited mobility are directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog number and lot number are unknown, however, the alleged celect is manufactured and inspected according to controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
Patient allegedly received an implant on (b)(6)2013 via the right internal jugular vein due to after being diagnosed [vt] deep vein thrombosis (dvt)/pulmonary embolism (pe).The patient alleges tilt and vena cava perforation.The patient further alleges chest pain, leg pain, pain throughout the body, and limited mobility.On (b)(6)2018, per a report from computed tomography; ¿findings: there is an ivc filter in place with the apex located 3mm from the right lateral wall of the ivc and 16.7mm from the left lateral wall.The tip of the filter is 7mm below the right renal vein and 2cm below the left renal vein.The 4 dominant struts perforate and extend beyond the outer wall of the ivc ranging from 2mm along the right lateral wall to 5mm along the medial posterior wall into the retroperitoneal fat.No fractured or bent fragments are noted.Impression: ivc filter location and tilt as detailed above.¿.
|