It was reported that, during a cori-assisted tka procedure, a camera disconnection error happened just before collecting the post-op baseline rom.They recovered by re-booting the system, re-calibrating, and navigating back to the post op collection screen.The procedure was completed as planned with cori and there was a delay of less than 30 minutes.There was no impact to the patient.No other complications were reported.
|
The real intelligence tracking camera, part number rob10025, serial (b)(6) and used for treatment, was not returned for evaluation.A relationship between the reported event and the device could not be established.A visual/functional inspection cannot be completed as no device was returned for evaluation.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.While the reported problem was not confirmed during evaluation, the most probable contributing factor(s) to the reported problem is loose/bad connection between console and camera.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.If the product associated with this event is returned or provided at a future date, this evaluation will be reopened for investigation.
|