BLUE BELT TECHNOLOGIES REAL INTELLIGENCE TRACKING CAMERA; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number ROB10025
Device Problem Intermittent Communication Failure (4038)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Event Description
It was reported that during a cori-assisted ukr procedure, there was a set off camera initializing error once at the beginning of the case and once at the end.For the beginning of case error, they hard power off the system and restarted the case.There was a delay of less than 30 minutes.No other complications were reported.
 
Manufacturer Narrative
H3, h6: the real intelligence tracking camera, part number rob10025, serial unknown and used for treatment, was not returned for evaluation.A relationship between the reported event and the device could not be established.A visual/functional inspection cannot be completed as no device was returned for evaluation.While all products meet required manufacturing specifications prior to release a serial number or lot number is required to link the device to a dhr or nc investigation, or field service report.A complaint history review for similar reported/confirmed complaints has identified prior events.A contributing factor could be software issues.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.If the product associated with this event is returned or provided at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE TRACKING CAMERA
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key10438575
MDR Text Key204040408
Report Number3010266064-2020-01733
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberROB10025
Was Device Available for Evaluation? No
Date Manufacturer Received09/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-