• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S TITAN OTR; INFLATABLE PENILE PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S TITAN OTR; INFLATABLE PENILE PROSTHESIS Back to Search Results
Model Number QSR8201400
Device Problems Leak/Splash (1354); Pumping Problem (3016)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/23/2020
Event Type  malfunction  
Manufacturer Narrative
Lot number: 1780088.The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release and no trends were noted.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Once our evaluation is complete, a follow-up report will be submitted.
 
Event Description
According to the available information, pump no longer working - only one cylinder.A leak in the tubing by the pump was reported.The device was revised.
 
Event Description
This follow-up mdr is created to document the additional event information received for record #(b)(6).According to the available information the inflatable penile prosthesis was implanted on (b)(6) 2009 and revised on (b)(6) 2020 due to malfunction.Additional information received from the clinical sales representative indicated: ¿there was a comment discussed in regards to an leak in the tubing by the pump.¿ the device was not received for evaluation.Without the benefit of analyzing the device quality cannot confirm any observations or comment on the condition of the product.If the device becomes available, or additional information is received, the complaint will be re-evaluated according to procedures.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database revealed no trends for this lot.Review of nonconforming reports revealed no nonconformances for this lot.No capas are associated with this lot.
 
Manufacturer Narrative
Based on the available information, the reported complaint is identified in the risk management documentation and reviewed routinely with management to monitor complaint trends as part of post market surveillance.No corrective action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN OTR
Type of Device
INFLATABLE PENILE PROSTHESIS
Manufacturer (Section D)
COLOPLAST A/S
holtedam 1
humlebæk 3050
DA  3050
MDR Report Key10438649
MDR Text Key203905169
Report Number2125050-2020-00761
Device Sequence Number1
Product Code FHW
Combination Product (y/n)N
PMA/PMN Number
P000006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberQSR8201400
Device Catalogue NumberQSR820
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/23/2020
Initial Date FDA Received08/21/2020
Supplement Dates Manufacturer Received07/23/2020
Supplement Dates FDA Received10/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age65 YR
-
-