| Model Number |
FRED4012-PMA |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
Thrombus (2101)
|
| Date of Event |
07/23/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was not returned to the manufacturer for evaluation.The root cause of the complaint cannot be determined.The instructions for use (ifu) identifies vessel stenosis or thrombosis as a potential complication associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the fred was deployed without incident in the internal carotid artery (ica).Post-deployment, thrombus developed on the proximal portion of the fred and the ica became occluded.Integrilin was administered intra-arterially and the ica opened up.It appeared that the inner stent of the fred was separated from the outer stent.The stent was opened and the patient woke up neurologically intact.There was no reported sequela.
|
| |
|