| Model Number |
00884450026571 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/27/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a diagnostic procedure, a catheter tip detached within the patient.The physician had acquired retrograde, arterial access and during catheter manipulations within the patient the 4f catheter tip detached.The physician successfully retrieved the foreign body from the patient with a vascular snare device without additional consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was returned for evaluation.The complaint was confirmed; however, the root cause could not be determined; however, it is likely that significant force was applied to the device during use.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
| |
|