ENDOLOGIX AFX; BIFURCATED STENT GRAFT Back to Search Results
Model Number BA25-100/I16-40
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Aneurysm (1708); Failure of Implant (1924)
Date of Event 07/29/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device remains implanted in the patient; therefore, it will not be returned for evaluation.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx with duraply.Device remains implanted in patient.
 
Event or Problem Description
The patient was initially implanted with an afx bifurcated stent graft, an afx vela suprarenal to treat an abdominal aortic aneurysm (aaa).Approximately 6 years post initial procedure, during a routine follow up, computerized tomography identified aneurysm enlargement.At re-intervention, a small type iiia endoleak was identified.The leak was resolved with an afx vela infrarenal bridging the overlap.The patient was released home.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident could not be completed.No medical records nor medical imaging relevant to the reported adverse event/incident was received by endologix.The user facility was unable to provide this information.Due to the absence of medical records and medical imaging; device, use, procedure, and/or anatomy relatedness to this adverse event/incident could not be evaluated.The final patient status remains unknown.The final patient status was not made available to endologix.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx with duraply.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
MDR Report Key10440200
Report Number2031527-2020-00243
Device Sequence Number19500840
Product Code MIH
UDI-Device Identifier00818009010032
UDI-Public(01)00818009010032(17)170611
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other,Recall
Type of Report Initial,Followup
Report Date (Section B) 07/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/11/2017
Device Model NumberBA25-100/I16-40
Device Lot Number1252293017
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/29/2020
Supplement Date Received by Manufacturer10/09/2020
Initial Report FDA Received Date08/21/2020
Supplement Report FDA Received Date10/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0006-2019
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX VELA SUPRARENAL (LN 1252387015)
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
-
-