| Model Number |
MMT-723LNAB |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Insulin pump passed the rewind, basic occlusion, prime, compromised force sensor system and displacement test.Pump received with stuck motor error alarm loop during bolus delivery.The motor was tested outside of the device and passed.The motor may have had and intermittent failure that was not detected during our testing.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had motor error alarm.Customer was able to successfully clear the alarm.Customer was able to complete pump rewind.No harm requiring medical intervention was reported.The insulin pump was returned for analysis.
|
| |
|
Search Alerts/Recalls
|