| Model Number |
5442138 |
| Medical Device Problem Codes |
Break (1069); Device Fell (4014)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from healthcare provider (hcp) via field representative regarding patient with spinal fracture involved in spinal reduction and internal fixation procedure.It was reported that during procedure, the fixation crosslink broke and fell off, that made unable to fix.There is no fragment left in the patient.There is no patient symptoms or complications reported as the result of event.There is no additional surgery performed as a result of this event.Patient is alive and no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary - visual and microscopic review confirms that the tip on the break off screws was fractured.A microscopic review of the fracture face reveals that the fracture is consistent with torsional overload with a bending moment introduced as the screw head pushed against the rod.If the screw is driven deeper using the inner hex after the break-off portion of the screw head has come off, there is a chance to overload the tip of the screw.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|