MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Model Number 5440030
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/03/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Evaluation of the returned device was not completed at the time of this report.A follow up report will be sent when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare provider via a manufacturer representative regarding a patient with pre-operative diagnosis of anterior lumbar 4,5 vertebral tuberculosis lesions.It was reported that intra-op,screw plug slipped.The thread of the screw plug was easy to wear, and the screw plug could not be locked tightly.Device is explanted and it will be returned.There were no patient symptoms or complications associated with this event.
 
Additional Manufacturer Narrative
H3: product analysis summary- visual and macroscopic inspection confirmed the threads of the break off screw have been damaged.The thread crest and flank damage appears to have initiated at the start of the thread, and is consistent around the damaged portion of the thread.Functional evaluation with a sample mas found the set screw is unable to be fully engaged in the mas head.This is consistent with misalignment of the mas and set screw threads during construct assembly.H6: fdm and fdr updated as per analysis results.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key10441524
Report Number1030489-2020-01131
Device Sequence Number2816852
Product Code KWP
UDI-Device Identifier00613994790705
UDI-Public00613994790705
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K102555
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5440030
Device Catalogue Number5440030
Device Lot NumberH5577881
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2020
Initial Date Received by Manufacturer 08/03/2020
Supplement Date Received by Manufacturer10/20/2020
Initial Report FDA Received Date08/24/2020
Supplement Report FDA Received Date11/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age23 YR
-
-