CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE Back to Search Results
Model Number DBP-125MICRO145
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Codes Vessel Or Plaque, Device Embedded In (1204); Vascular Dissection (3160)
Date of Event 07/30/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements prior to distribution.(b)(4).
 
Event or Problem Description
During use of a diamondback peripheral orbital atherectomy device (oad) in the anterior tibial artery, the device interacted with tissue will spinning and became stuck to the vessel wall.The oad was turned off and on multiple times to release the crown, and was removed from the patient.Tissue was noted around the device crown and per the opinion of the physician, a non-visible dissection may have been present prior to the procedure which resulted in the issue.A type b dissection was present following removal of the oad and was resolved with a long balloon inflation.The vessel was then re-wired and another oad was used to complete the procedure.The patient was stable following the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old highway 8 nw
st. paul, mn, mn
MDR Report Key10442045
Report Number3004742232-2020-00256
Device Sequence Number16757407
Product Code MCW
UDI-Device Identifier10850000491172
UDI-Public(01)10850000491172(17)220228(10)309977
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K190634
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2022
Device Model NumberDBP-125MICRO145
Device Catalogue Number7-10057-01
Device Lot Number309977
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/31/2020
Initial Report FDA Received Date08/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient Weight116
-
-