| Model Number |
DBP-125MICRO145 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Codes |
Vessel Or Plaque, Device Embedded In (1204); Vascular Dissection (3160)
|
| Date of Event |
07/30/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements prior to distribution.(b)(4).
|
| |
|
Event or Problem Description
|
|
During use of a diamondback peripheral orbital atherectomy device (oad) in the anterior tibial artery, the device interacted with tissue will spinning and became stuck to the vessel wall.The oad was turned off and on multiple times to release the crown, and was removed from the patient.Tissue was noted around the device crown and per the opinion of the physician, a non-visible dissection may have been present prior to the procedure which resulted in the issue.A type b dissection was present following removal of the oad and was resolved with a long balloon inflation.The vessel was then re-wired and another oad was used to complete the procedure.The patient was stable following the procedure.
|
| |
|