Model Number N/A |
Event Date 07/31/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source: (b)(6).Product is in process of being returned to zimmer biomet for the investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2020-03299.
|
|
Event Description
|
It was reported that during incoming inspection at the warehouse, debris was found in the sterile packaging.There was no patient involvement.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
Complaint sample was returned and reviewed against the reported event.Evaluation of the returned product/photographs provided confirmed foreign debris is present inside the sterile packaging, and the sterile packaging remains sealed.The reported event is confirmed.Dhr was reviewed and no discrepancies were found.The likely condition of the device when it left zimmer biomet is non-conforming to specification.The root cause of the reported event is the operator not following the work instructions provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional event information to report at this time.
|
|