• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION 2.0 LACTOSORB SYSTEM 2.0 X 5MM DIRECT DRIVE SCREW; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION 2.0 LACTOSORB SYSTEM 2.0 X 5MM DIRECT DRIVE SCREW; SCREW, FIXATION, BONE Back to Search Results
Model Number N/A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/07/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet (b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2020-00384, 0001032347-2020-00385.Medical products: unknown blade catalog#: ni lot#: ni; 2.0 lactosorb system 2.0 x 7 mm direct drive screw, part# 915-2201, lot# 775440; 2.0 lactosorb system 2.0 x 5 mm direct drive screw, part# 915-2200, lot# 479600.The user facility is foreign; therefore, a facility medwatch report will not be available.Report source - (b)(6).
 
Event or Problem Description
It was reported that during an oral surgery, the surgeon was unable to grasp the screw with the driver and the screw fell off.No adverse events have been reported as a result of the malfunction.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d4, d10, g4, g7, h2, h3, h6, h10.The customer returned qty one 915-2200 lot 849900 lactosorb screws for evaluation due to lack of retention complaints.A visual inspection of the screws showed light signs of attempted use.For further analysis the screws were returned to warsaw for further investigation by ops manufacturing technician.The reported testing results show that the returned product was all found to fail the g01355 no-go test has been deemed out of spec.The complaint is confirmed.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.The root cause of the reported issue is attributed to issues in the manufacturing process leading to undersized parts being manufactured.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0 LACTOSORB SYSTEM 2.0 X 5MM DIRECT DRIVE SCREW
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key10442243
Report Number0001032347-2020-00383
Device Sequence Number3975148
Product Code HWC
UDI-Device Identifier00841036054687
UDI-Public00841036054687
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K011139
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/17/2024
Device Model NumberN/A
Device Catalogue Number915-2200
Device Lot Number849900
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/12/2020
Supplement Date Received by Manufacturer01/22/2021
Initial Report FDA Received Date08/24/2020
Supplement Report FDA Received Date02/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-