• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1550250-38
Medical Device Problem Codes Difficult to Remove (1528); Difficult to Advance (2920); Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a left anterior descending (lad) artery.A 2.5x38mm xience sierra was advancing when resistance was met with an unspecified guiding catheter.Resistance was also felt with the unspecified guiding catheter during removal and once the device was removed it was noted that the distal edge of the stent was not crimped on properly to the balloon.Another same size stent was used to complete the procedure.There were no adverse patient effects and no clinically significant delay.No additional information was provided.
 
Additional Manufacturer Narrative
A visual, functional and dimensional inspection was performed on the returned device.The reported material deformation was confirmed.The reported difficult to advance/position and difficult to remove were not confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported difficulties.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10442250
Report Number2024168-2020-07030
Device Sequence Number2713990
Product Code NIQ
UDI-Device Identifier08717648227141
UDI-Public08717648227141
Combination Product (Y/N)Y
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/09/2020
Device Model Number1550250-38
Device Catalogue Number1550250-38
Device Lot Number9100341
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2020
Initial Date Received by Manufacturer 07/30/2020
Supplement Date Received by Manufacturer09/01/2020
Initial Report FDA Received Date08/24/2020
Supplement Report FDA Received Date09/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-