• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. BIPOLAR HIP; SHELL, BIPOLAR, MODULAR, ENCORE, SIZE46

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. BIPOLAR HIP; SHELL, BIPOLAR, MODULAR, ENCORE, SIZE46 Back to Search Results
Model Number 412-02-046
Health Effect - Clinical Code No Code Available (3191)
Date of Event 07/27/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery due to instability.The previous surgery and the surgery detailed in this event occurred 60 days apart.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to instability.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery - patient has a djo hemi-arthroplasty that surgeon revised for instability, she was done out of state.He anticipate going up on neck length and decreasing the size of head 1-2 mm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPOLAR HIP
Common Device Name
SHELL, BIPOLAR, MODULAR, ENCORE, SIZE46
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445,
MDR Report Key10442768
Report Number1644408-2020-00683
Device Sequence Number16848215
Product Code KWY
UDI-Device Identifier00888912077378
UDI-Public(01)00888912077378
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K953510
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/30/2024
Device Model Number412-02-046
Device Catalogue Number412-02-046
Device Lot Number991B1173
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2020
Initial Report FDA Received Date08/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
497-28-035 LOT 631C1338
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-