• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC MINIMAL ACCESS ATTACHMENT, STRAIGHT; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC MINIMAL ACCESS ATTACHMENT, STRAIGHT; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Model Number MA-DRIVER
Medical Device Problem Codes Vibration (1674); Excessive Heating (4030)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.However, since the investigation is still on-going, the assignable root cause could not be determined at this time.Once investigation has been completed, a supplemental medwatch will be submitted accordingly.(b)(4).
 
Event or Problem Description
It was reported by (b)(6) that during service and evaluation, it was determined that the attachment device generated heat and strong ¿heavy¿ vibrations.It was further determined that the device failed pretest for max temperature of ma attachment.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2020.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: correction: device availability: d10: the device availability date was incorrect in the initial report.The date has been updated from (b)(6) 2020 to (b)(6) 2020.Device evaluation: this device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, an evaluation was performed and it was determined that the warning triangle was missing, and the labeling was illegible and etching.It was further observed that the device generated heat and vibration.It was further determined that the device failed pretest for max temperature of ma attachment and visual assessment (verify that all markings are present and legible).Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to component failure from normal wear.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINIMAL ACCESS ATTACHMENT, STRAIGHT
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
MDR Report Key10442817
Report Number1045834-2020-01360
Device Sequence Number14457091
Product Code HBE
UDI-Device Identifier00845384004370
UDI-Public00845384004370
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K042783
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMA-DRIVER
Device Catalogue NumberMA-DRIVER
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2020
Initial Date Received by Manufacturer 08/10/2020
Supplement Date Received by Manufacturer09/10/2020
Initial Report FDA Received Date08/24/2020
Supplement Report FDA Received Date09/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-